Annual service, calibration and validation of Millipore ultra pure water system

Annual service, calibration and validation of our clients Millipore ultra pure laboratory water system, Cryomed supplied and replaced all consumables and calibrated the ultra pure water to 18.2 megaohms

Panasonic -150 chest freezer 2nd stage system refrigerant leak

IMG_7443IMG_7443 Refrigerant Leak on the 2nd stage of a Panasonic -150 ULT freezer model MDF-C2156VAN

Beckman centrifuge refrigeration system repair

IMG_7290

BOOST PUMP REPLACEMENT TO SUEZ WATER PURIFICATION SYSTEM

Removal and replacement of boost pump from our clients pure water Suez laboratory water system after our engineers recent pump failure diagnosis, system now fully operational with a water purity of 18.2 mega ohms.

Suez laboratory water leak

IMG_6816 IMG_6816 Client contacted Cryomed scientific when a leak appeared on their suez laboratory ultra pure water system , our engineer quickly pinpointed the issue, removed and replaced leaking fitting and had the system fully back operational within the hour.

Ultra pure water system service and validation

Annual service, calibration and validation of our clients triple red laboratory ultra pure water system.

IQ,OQ,PQ Validation of a new research facility

Panasonic blood fridge service and calibration

6 monthly service and calibration of our clients Panasonic blood fridge chart recorder and temperature display of upper and lower temperature readings

PHCBI Co2 incubator with co2 leakage fault

Customer called Cryomed scientific when they discovered their phcbi co2 incubators using excessive amounts of co2 , on arrival our incubator specialist engineer quickly diagnosed the fault and rectified the faults and having the incubators fully operational on the first visit

Helmer cooled incubator service and calibration

6 monthly service and calibration of our clients Helmer cooled incubator

Eppendorf laboratory co2 incubator fault

Client reported to Cryomed scientific an issue with their Eppendorf co2 incubator, Cryomed scientifics engineer quickly diagnosed the fault as the incubator control sensor, once the new temperature sensor was received from Eppendorf the sensor was removed and replaced, calibrated and the Eppendorf co2 incubator was back fully operational

Rephile RO Water purification system installation

Cryomed scientific supplied and installed a new Raphile laboratory RO water system to our clients laboratory, installed, calibrated and fully operational in 4 hours

Elga laboratory ultra pure water system service

New client contacted Cryomed scientific when their Elga ultra pure laboratory water system wasn’t achieving 18.2 mega ohms, our engineer quickly diagnosed that the RO membranes required replacing and the system required a full service and calibration , RO membranes replaced and annual service of the Elga ultra pure water system left the system fully… Continue reading Elga laboratory ultra pure water system service

Panasonic -150 Freezer leak and pressure test of 2nd stage system

A new client contacted Cryomed scientific when they found their Panasonic -150 ultra low freezer had failed, We uplifted the freezer back to our workshop when our engineer discovered the 2nd stage system of the Panasonic-150 freezer to have lost all its refrigerant charge. Once the Panasonic -150 freezer arrived at the workshop our engineers… Continue reading Panasonic -150 Freezer leak and pressure test of 2nd stage system

Statebourne liquid nitrogen system servicing

Annual servicing of our clients Statebourne liquid nitrogen systems , fully calibrated and serviced within one day

Hoshizaki laboratory Ice flaker repairs

Customer contacted Cryomed scientific to resolve issues with 2 hoshizaki laboratory ice flaker machines in their labs, both hoshizaki machines were in desperate need of a full service and replacement condenser fan motors , both motors replaced and both hoshizaki laboratory ice flaker machines serviced on the first visit and fully back in operation, happy… Continue reading Hoshizaki laboratory Ice flaker repairs

Infors shaker refrigerated incubators

Repairs of our customers refrigeration system on their infors refrigerated shaker incubator , also our engineer replaced the platelet actuator arms

Eppendorf -80 ult freezer door vacuum release issue

Customer called Cryomed scientific when they couldn’t open the outer freezer door to access their samples due to the door being jammed closed, Our engineer quickly diagnosed the issue as being an ice build up on the inner door over the vacuum release system, once the ice was removed the door was now easily opened… Continue reading Eppendorf -80 ult freezer door vacuum release issue

Purite ultra pure laboratory water system leak

IMG_5870 Client called Cryomed scientific when their Purite water purification system starting leaking , on closer investigation from our engineer he traced the leak to a faulty inlet water pressure guage, our water specialist quickly removed and replaced the leaking water pressure guage with a new one and the Purite ultra pure water system was… Continue reading Purite ultra pure laboratory water system leak

HAIER AND LABCOLD ULT -80 FREEZER SERVICING AND CALIBRATION

Annual service and calibration for our clients Haier and Labcold -80 ultra low temperature freezers

Millipore Ultra pure Laboratory water system fault

A call from our client regards flow issues with their Millipore laboratory ultra pure water system , our engineer quickly diagnosed the issue as being a failure of the flow meter on the Millipore system, removal and replacement of the flow meter quickly resolved the issue and the system was placed back into full operation.

Hot Room over temperature issue

Our client called Cryomed scientific to investigate an issue with over heating on a constant temperature hot room, on investigation of the temperature controller our engineer found the controller to be faulty and operating the heaters at 2 degrees above set point, After close discussions with our client our engineer removed and replaced the temperature… Continue reading Hot Room over temperature issue

Triple Red Avidity Science ultra pure water system fault

client called Cryomed scientific to report quality issues with the triple red Avidity science ultra pure water system, our engineer quickly diagnosed the fault as the quality sensor who then removed and replaced the sensor, calibrated the water quality by using our ultra pure water quality meter and retuned system to service

Coldroom compressor change

Customer complained that there was an unusual noise coming from the condensing unit that serves several coldrooms , our engineer quickly diagnosed the fault as the electronic Copeland compressor was running but not cooling, within 48 hours we had quoted the client , ordered and installed the new compressor with minimum down time to our… Continue reading Coldroom compressor change

CRYOMED SCIENTIFIC UKAS ISO 9001 ACCREDITATION

Cryomed Scientific are please to announce their annual ISO9001 accreditation has been awarded for 2024 with no non conformances , well done to all staff and Accendo consultants who perform our external audits.

Heraeus co2 incubator fan motor replacement

Customer contacted Cryomed Scientific when their Heraeus co2 incubator started fluctuating the co2 levels on the display and alarming, our engineer quickly assessed the issue as the internal fan motor seizure, after discussions with the department we then ordered and replaced the fan motor , fully calibrated co2 and temperature displays and left the incubator… Continue reading Heraeus co2 incubator fan motor replacement

Nuaire co2 incubators repaired,serviced and calibrated

Customer decided to move from Avidity to Cryomed Scientific to repair, service and calibrate their co2 incubators. Removal and replacement of air pump cured their issue with irratic c02 levels on one incubator,both incubators had their air, co2 and hepa filters removed and replaced, calibration performed by our Gemtech analyser, happy customer and over £2,000… Continue reading Nuaire co2 incubators repaired,serviced and calibrated

New Brunswick Scientific ULT -80 freezer servicing and calibration

Annual service and calibration of Cryomed scientific’s client New Brunswick -80 upright freezers, our engineer identified 2 faulty alarm batteries and a faulty control board which was causing the compressors to short cycle intermittently, not an easy fault find.IMG_5282

Ultra low temperature -80 freezers serviced and calibrated

Annual service and calibration of our clients 13 ultra low temperature -80 freezers

Panasonic ULT freezer 2nd stage compressor replacement

Panasonic -80 freezer 2nd stage compressor removal and replacement, full vacuum and pressure test of system and back down to -80 and ready for use once again within 24 hours

Fisher Scientific Basic co2 Incubator installation and service

Installation and Calibration of a new Fisher basic co2 incubator for our client

PANASONIC -150 ULTRA LOW TEMPERATURE FREEZER FAN MOTOR REPLACEMENT AND TIMER RESET

Customer called Cryomed scientific when there was a noise coming from their Panasonic -150 freezer and an alarm was indicated, our engineer quickly diagnosed the fault as a faulty condenser fan motor which was removed and replaced within the hour, master reset of alarm timers and freezer was back fully operational and operating at -150… Continue reading PANASONIC -150 ULTRA LOW TEMPERATURE FREEZER FAN MOTOR REPLACEMENT AND TIMER RESET

Panasonic co2 incubator leak

Customer called Cryomed scientific when they realised they were using too much co2 and the co2 incubator wasn’t achieving set point of 5% Our engineer quickly diagnosed and located the faulty burst tubing in the back of the co2 incubator, repaired and customer was back operational within a couple of hours.

Panasonic ult chest freezer compressor failure

Thermo scientific replacement ultra pure filter on our clients laboratory water purification system

Infors cooled incubator shaker repair

Removal and replacement of internal air circulating fan motors and control board on our clients infors cooled incubator shaker system

BINDER CO2 INCUBATOR FAULTS

IMG_4947A new client of Cryomed scientific contracted our service and validation department to install and commission their newly acquired binder co2 incubator, once fully operational and found to have passed our onsite calibration it was found the temperature controller was faulty and causing the heater system to keep heating until the safety switch activated and… Continue reading BINDER CO2 INCUBATOR FAULTS

New Brunswick ult freezer failure

SANYO ULT freezer 2nd stage system choking

Customer reported freezer fluctuating in temperature, on investigation by our ULT freezer engineer it was quickly diagnosed the 2nd stage refrigeration system was partially choking, removal and installation of a new filter drier on the 2nd stage refrigeration system cured the fault, freezer fully back operational and another happy customer.

Panasonic Blood fridge service and calibration

Fan motor replacement on Heraeus -80 chest freezer

Customer called Cryomed scientific due to thei Heraeus ultra low temperature freezer overheating, our engineer quickly identified the condenser fan motor to be faulty, due to our large stock of ult freezer parts our engineer replaced the fan motor and had the freezer fully operational and back down to -80 within the hour.

Avidity laboratory water system pump replacement

Customer contacted Cryomed scientific when they had a water leak from their Avidity science water purification system, our engineer quickly diagnosed the fault as the pump head leaking, pump was removed and replaced with system back fully operational.

Eppendorf replacement condenser fan motor

A new customer contacted Cryomed scientific when their Eppendorf -80 freezer broke down , our engineer attended site and replaced the condenser fan motor on the 1st visit as we carry a fully comprehensive parts list for all -80 freezers, freezer fully operational and back down to -80 before the end of the day, ready… Continue reading Eppendorf replacement condenser fan motor

Avidity water purification service and calibration

Annual service, calibration and validation of our customers Avidity ultra pure laboratory water purification system, all consumables supplied by Cryomed scientific, 18.2 m/ohms calibration completed with our calibration meter, temperature and time calibrated with our UKAS certified equipment.

Service and calibration of Binder co2 and O2 controlled incubator

Annual service and calibration of Binder co2, oxygen level and temperature control levels

Water pump repair to Elga ring main water purification system

University customer called Cryomed scientific when their Elga water purification ring main system stopped filling the reservoir tank and the system was leaking, Within an hour our engineer had removed the pump , repaired the leak and freed the seized pump, system back fully operation with minimum downtime for our client

RS BIOTECH EX DEMO ULT FREEZER DELIVERY

Another delivery and installation of our stock ult freezers, This one is an rs biotech badged New Brunswick-80 upright freezer that was purchased by Newcastle university just before the Christmas break, All our ex-demo and refurbished-80 freezers come complete with 12 month parts and labour warranty, fully serviced, calibration certification and temperature mapping if requested.… Continue reading RS BIOTECH EX DEMO ULT FREEZER DELIVERY

Used ult freezers for sale

Cryomed scientific have now taken stock of ult -86 freezers and these shall appear on our web site for sale at massively reduced prices from new, Featuring all the freezers manufacturers such as EPPINDORF,THERMO,PHCBI,SANYO,PANASONIC All freezers are supplied with 12 months parts and labour warranty, free delivery, calibration certificate,24 hour temperature map as standard. All… Continue reading Used ult freezers for sale

LEEC WARMING CABINET REPAIR

Customer contacted Cryomed scientific when two of her LEEC warming cabinets stopped operating and over heating,after diagnosing the fault on the phone to our customer as faulty control thermostats we dispatched our engineer,once on site both thermostats were replaced at the same visit due to a stock of parts for the LEEC warming cabinets, Within… Continue reading LEEC WARMING CABINET REPAIR

Annual service,calibration and validation of Millipore ultra pure integral 15 water purification system with A10 UV

An Annual service, calibration and validation of our clients ultra pure laboratory water system Millipore integral 15 with A10 uv Removal and replacement of all consumables supplied by Cryomed scientific cheaper than Millipore , replacement of RO membranes complete sanatisation of tank , all completed in one visit and customers system back fully operational operating… Continue reading Annual service,calibration and validation of Millipore ultra pure integral 15 water purification system with A10 UV

Annual service and Calibration of Leica cryostat

Customer contacted Cryomed scientific when their cryostat started showing ‘help’ on display, after servicing , calibration and reset of the help display cryostat is now back fully operational and operating at -25

Triple red’s SG ultra pure laboratory water system annual service and calibration to 18.2 m /ohms

Annual service and calibration of our customers Triple red / SG Ultra pure laboratory water purification system calibrated to 18.2 m/ohms, sanitisation of water storage tank and flushed to drain in the same day , fully back operational for our customer to use

Elga laboratory water purification system annual service and calibration to 18.2 m/ohms

The annual service and calibration of our customers Elga centra 200 laboratory ring main water purification system, consumables all changed sanitised and flushed to drain all within the same day , fully operational again for our customer to use with water quality at 18.2 m/ohms

-86 Ultra Low Temperature freezer servicing

An annual Panasonic and Gram bioline -80 freezer servicing and calibration, mobile ult freezer .

Avidity water system repair

Customer called Cryomed scientific when the manufacturer was unable to locate and repair water leak on the pure water system, within the hour our engineer located and replaced leaking part, System back fully operational and customer has asked for quotation to contract Cryomed scientific with all their water purification systems.

Annual -80 freezer servicing

Annual -80 freezer serving and calibration on these Eppendorf and Panasonic upright -80 freezers for our customer in Newcastle university

CBS PANASONIC LN2 VALIDATION

Cryomed scientific was contracted to perform the installation qualification and operational validation for their liquid nitrogen storage facility bank , 4 Panasonic CBS liquid nitrogen storage vessels validated and ready for performance validation in two days.

Oil separator removed and replaced on 2nd stage system of Thermo -80 ult freezer

New customer asked Cryomed scientific to diagnose the intermittent fault on their Thermo -80 upright ult freezer, freezer operating at -80 on arrival but after looking at temperature graphs it was easy to diagnose the intermittent failure of the 2nd stage refrigeration systems oil separator, Post ordering a replacement oil separator our engineer removed faulty… Continue reading Oil separator removed and replaced on 2nd stage system of Thermo -80 ult freezer

Annual service and calibration of laboratory fridges and freezers

Pleased to perform another annual service and calibration to our long standing client in edinburgh, Fridges and freezers serviced and single point calibration to traceable UKAS standards

-80 ULTRA LOW TEMPERATURE PANASONIC AND EPPENDORF FREEZER SERVICING AND CALIBRATION

Annual calibration and servicing of 2 -86 ult freezers for our client , Panasonic ult freezer and Eppendorf ult freezer, full calibration of the 20 comark temperature sensors of their monitoring system

Avidity Laboratory water system service and calibration

Annual service and calibration of Avidity ultra pure laboratory water system, consumables all replaced and timers reset, back in operation within a couple of hours, calibration completed to national standards

Coldroom repair

Customer contacted Cryomed scientific when their +4 walk-in coldroom failed, our engineer quickly diagnosed the system to have lost all its refrigerant charge, following a pressure test of the pipework a leak was found and repaired on the expansion valve flare nut, after vacuuming the system and fully charging the system with refrigerant the customer… Continue reading Coldroom repair

VINDON Climatic chamber repair

Customer contacted Cryomed scientific when their maintenance and validation company couldn’t repair their Vindon climatic chamber, Leak and pressure test of the refrigeration system identified a leak on the suction valve once the valve was fully tightened and repaired another leak and pressure test followed, fully charging the system with refrigerant, once the system fully… Continue reading VINDON Climatic chamber repair

Panasonic -80 ULT FREEZER COMPRESSOR REPLACEMENT

Removal and replacement of 2nd stage compressor on our customers Panasonic -80 upright ULT Freezer, Repair performed on site and fully operational by the end of the day, pressure tested, leak tested and fully charged with 1st and 2nd stage refrigerants, Saves customer time and money moving freezer off site,

Eppendorf Cryocube -80 upright freezer repair

New condenser fan replacement on Eppendorf Cryocube -80 upright freezer

Panasonic co2 incubator fault

Customer contacted Cryomed scientific to investigate co2 leaking from incubator and alarming, Attended site same day and our engineer found the incubator co2 inlet valve leaking , having a valve in stock we replaced valve and filters to have the incubator back fully operational within the hour

Elga water system service and validation

Annual service and validation of our customers Elga laboratory water purification system , consumables replaced and sanitisation of system, back in operation the same day.

Out of hours callout to Eppendorf Cryocube -80 Freezer

Customer calls Cryomed scientific 24 hour emergency callout hotline to attend their failed freezer at 10.00pm, a quick diagnostic of the fault by our engineer of the fan failure, temporary fan installed to keep freezer fully operational at -80 until new fan ordered, happy customer with samples all safe and secure.

Sanyo -80 ultra low temperature freezer fault

A new customer contacted Cryomed scientific to diagnose a fault with their Sanyo -80 ultra low temperature freezer, our engineer found the 2nd stage compressor seized and blowing fuse on start up. Quotation send to customer to perform the repair on site and return their freezer to -80

Thermo Revco ultra low -140 chest freezer service

Annual service of our clients Thermo Revco -140 freezer

Panasonic -80 upright ult freezer replacement 2nd stage compressor and oil sep

Replacement of 2nd stage compressor and oil separator on our clients Panasonic -80 ult freezer, total job performed and completed on customers premises ,our competitors usually remove from site to repair costing the end user more expense to have uplifted and delivered back and to be out of use for weeks instead of days,

Panasonic -86 ULT FREEZER FAULT DIAGNOSE

 Customer contacted Cryomed scientific to investigate intermittent issue with their Panasonic ultra low temperature freezer. The customer reported intermittent fault of temperature rising then dropping back down to temperature on a daily basis, Eventually our engineer traced the fault to the 2nd stage refrigeration system, new compressor and oil separator required, quote and go… Continue reading Panasonic -86 ULT FREEZER FAULT DIAGNOSE

Millipore essential 24 laboratory water purification system RO membrane and boost pump replacement

Customer in Southampton contracted Cryomed scientific to replace RO membranes and boost pump on their Millipore essential 24 water purification system, their operating quality is now 3m/s instead of 60m/s and now delivering 25 litres/hr instead of 6 litres/hr

Fisher Accu20 annual service

Nottingham university invited Cryomed to service their Fisher Accu20 water purification system, replaced all consumables and flushed system through, left system fully operational with water quality at 18.2 m/ohms

Labcold cooled incubator repair

Customer complained that the system chart was irratic in temperature recording and called Cryomed in to investigate, we found that the refrigeration system to have lost some of its refrigerant charge due to a leak on the charging schrader, we replaced schrader and vacuumed system, fully charged with refrigerant and now chart is operating at… Continue reading Labcold cooled incubator repair

Thermo Haake -50 dip cooler repair

Customer contacted Cryomed to repair their Haake Thermo -50 dip cooler , after leak tested both refrigeration systems , leaks were found and repaired, both systems vacuumed and re charged with relevant refrigerants , running the system down to -50 , cascade refrigeration system working perfectly

Edwards freeze dryer refrigeration repair

Customer in Edinburgh University contacted Cryomed to repair the refrigeration system on their 30 year old Edwards freeze dryer, refrigeration system required retrofitted to operate on a more suitable refrigerant to achieve -40, delighted customer when we achieved -42 under vacuum, another piece of equipment saved from being made redundant.

Thermo forma ult -86 upright freezer repair

New Customer contact Cryomed to investigate Thermo -80 freezer tripping circuit breaker , investigated issue and found faulty wiring tripping breaker on -80 Thermo freezer, repaired wiring and ult -80 freezer back down to set point of -85 in one visit, Happy customer that didn’t need to move any of their samples and reduce down… Continue reading Thermo forma ult -86 upright freezer repair

Elga laboratory pure water ring main system service

Annual service and calibration of this Elga laboratory pure water ring main system, saving customer over £1,000 on manufacturers costs for service and water consumable filters

Panasonic -80 ult chest freezer repair, control board replacement

Customer contacted Cryomed when over the weekend their Panasonic -80 ult freezer failed, we attended site and identified that the main control board was not sending a signal out to run 2nd stage compressor, we removed and replaced control board on the Panasonic -80 chest freezer , freezer was back down to -80 by the… Continue reading Panasonic -80 ult chest freezer repair, control board replacement

Panasonic co2 incubator gas leakage

Advised by customer hissing sound from Panasonic co2 incubator and losing co2 from bottle Leak tested system and found inlet co2 valve leaking, Due to Cryomed carrying spares for Panasonic co2 incubators we managed to remove and replace co2 valve in one visit and have the Panasonic incubator back fully operational for our happy customer.

-80 freezer repair and fan motor replacement

Customer requested fan motor changes to their -80 ult freezers, Fans replaced and alarm lights reset, no need for moving samples, freezer temperatures kept -80 throughout repairs, happy customer with a saving of over £1,000 on manufacturers repair.

Thermo forma co2 incubator service and repair

Customer contacted Cryomed to service and repair 3 of 4 Thermo co2 incubators, due to being over 20 years old we had to use parts from incubators to make 2 out of three operational, full calibration of temperature, co2 and humidity, new hepa filters , co2 and air filters installed

Eppendorf cooled centrifuge repair

IMG_0538 Customer reported Eppendorf cooled centrifuge not cooling, Identified leak on system, repaired on site, fully charged system with refrigerant and oil leaking the centrifuge fully operational on the first visit

New Brunswick -80 freezer 2nd stage compressor replacement

Customer reported New Brunswick scientific -80 freezer blowing fuses, Identified 2nd compressor faulty, removed and replaced compressor back up and running within 24hours and down to -80.

Validation and service of Purite Laboratory water system

Contacted by a new client to supply consumables, service and validate their Purite laboratory water purification system Service and validation completed with full documentation and consumables on time and saving our new client over £1,500 on their previous supplier

Sanyo ULT Freezer fan motor replacement

Customer reported temperature fluctuating on their Sanyo ultra low temperature freezer, identified the fault as a seized condenser fan motor , due to our high stock levels for all ult freezers we quickly removed and replaced fan motor in one visit leaving the freezer operating at -80.

Service ,calibration and condenser fan motor replacement

Removed and replaced condenser fan motors on Panasonic -150 freezers for Newcastle University customer

Service and replacement boost pump on Suez Purite laboratory water purification system

Newcastle university approached ourselves to service and replace boost pump on Purite water purification system when the manufacturer quoted , Cryomed scientific supplied the required consumables,serviced, calibrated temperature and water quality sensors and replace boost pump at half the cost quoted to the customer from Suez.

Avidity Pure water system service

University of Newcastle contacted Cryomed to supply consumables and service their Avidity water purification system , we supplied all the consumables ,serviced and calibrated quality sensors, saving the customer over 30% in cost than using the manufacturer.