Sg/triple red/evoqua ultra pure water systems service and calibration

Annual service and calibration of our clients ultra pure laboratory water systems,consumables replaced including uv lamps, timers reset and calibrated using our ultra pure ukas accredited resistivity meter.

Ultra pure water system emergency callout Elga and triple red

Client called Cryomed when their Elga ultra pure water ringmain system failed, diagnosing the fault to be a faulty uv starting board issue, this was replaced at the same visit and back fully operational the same day. Client also had an issue with a triple red pure water system that was having an issue with… Continue reading Ultra pure water system emergency callout Elga and triple red

Purite laboratory water system repairs

IMG_8914 A couple of repairs on our clients Purite ultra pure water systems, back up and fully operational within 2 hours

Laboratory water ringmain service and calibration

Annual service and calibration of our clients ultra pure water ring main system, sanitisation of the ring main pipework and service of the Elga system with uv lamps

Ringmain Laboratory ultra pure water system service, sanitisation and calibration

Annual service, sanitisation and calibration for our clients ringmain ultra pure water system, consumables supplied and installed by Cryomed scientific, calibration of clients resistivity meter and alarm system performed with UKAS resistivity meter, Work completed and back fully operational in one day.

Annual service, calibration and validation of Millipore ultra pure water system

Annual service, calibration and validation of our clients Millipore ultra pure laboratory water system, Cryomed supplied and replaced all consumables and calibrated the ultra pure water to 18.2 megaohms

BOOST PUMP REPLACEMENT TO SUEZ WATER PURIFICATION SYSTEM

Removal and replacement of boost pump from our clients pure water Suez laboratory water system after our engineers recent pump failure diagnosis, system now fully operational with a water purity of 18.2 mega ohms.

Ultra pure water system service and validation

Annual service, calibration and validation of our clients triple red laboratory ultra pure water system.

Elga laboratory ultra pure water system service

New client contacted Cryomed scientific when their Elga ultra pure laboratory water system wasn’t achieving 18.2 mega ohms, our engineer quickly diagnosed that the RO membranes required replacing and the system required a full service and calibration , RO membranes replaced and annual service of the Elga ultra pure water system left the system fully… Continue reading Elga laboratory ultra pure water system service

Purite ultra pure laboratory water system leak

IMG_5870 Client called Cryomed scientific when their Purite water purification system starting leaking , on closer investigation from our engineer he traced the leak to a faulty inlet water pressure guage, our water specialist quickly removed and replaced the leaking water pressure guage with a new one and the Purite ultra pure water system was… Continue reading Purite ultra pure laboratory water system leak

Triple Red Avidity Science ultra pure water system fault

client called Cryomed scientific to report quality issues with the triple red Avidity science ultra pure water system, our engineer quickly diagnosed the fault as the quality sensor who then removed and replaced the sensor, calibrated the water quality by using our ultra pure water quality meter and retuned system to service

Avidity water purification service and calibration

Annual service, calibration and validation of our customers Avidity ultra pure laboratory water purification system, all consumables supplied by Cryomed scientific, 18.2 m/ohms calibration completed with our calibration meter, temperature and time calibrated with our UKAS certified equipment.

Water pump repair to Elga ring main water purification system

University customer called Cryomed scientific when their Elga water purification ring main system stopped filling the reservoir tank and the system was leaking, Within an hour our engineer had removed the pump , repaired the leak and freed the seized pump, system back fully operation with minimum downtime for our client

Triple red’s SG ultra pure laboratory water system annual service and calibration to 18.2 m /ohms

Annual service and calibration of our customers Triple red / SG Ultra pure laboratory water purification system calibrated to 18.2 m/ohms, sanitisation of water storage tank and flushed to drain in the same day , fully back operational for our customer to use

Elga laboratory water purification system annual service and calibration to 18.2 m/ohms

The annual service and calibration of our customers Elga centra 200 laboratory ring main water purification system, consumables all changed sanitised and flushed to drain all within the same day , fully operational again for our customer to use with water quality at 18.2 m/ohms

Avidity water system repair

Customer called Cryomed scientific when the manufacturer was unable to locate and repair water leak on the pure water system, within the hour our engineer located and replaced leaking part, System back fully operational and customer has asked for quotation to contract Cryomed scientific with all their water purification systems.

Avidity Laboratory water system service and calibration

Annual service and calibration of Avidity ultra pure laboratory water system, consumables all replaced and timers reset, back in operation within a couple of hours, calibration completed to national standards

Fisher Accu20 annual service

Nottingham university invited Cryomed to service their Fisher Accu20 water purification system, replaced all consumables and flushed system through, left system fully operational with water quality at 18.2 m/ohms