Sanyo ult freezer repair

Cryomed scientific welcomes A new client from Nottingham science park ,our engineer diagnosed and repaired the Sanyo ult freezer 2nd stage system being low in refrigerant ,within the same visit the Sanyo ultra low temperature -80 freezer was repaired and back down to set point of -80

Annual -80 ULT Freezer service

Annual service and calibration to our clients 17 ultra low temperature freezers including Sanyo,Phcbi,New Brunswick , Haier and Panasonic

SANYO ULT freezer 2nd stage system choking

Customer reported freezer fluctuating in temperature, on investigation by our ULT freezer engineer it was quickly diagnosed the 2nd stage refrigeration system was partially choking, removal and installation of a new filter drier on the 2nd stage refrigeration system cured the fault, freezer fully back operational and another happy customer.

Sanyo -80 ultra low temperature freezer fault

A new customer contacted Cryomed scientific to diagnose a fault with their Sanyo -80 ultra low temperature freezer, our engineer found the 2nd stage compressor seized and blowing fuse on start up. Quotation send to customer to perform the repair on site and return their freezer to -80