Annual service,calibration and validation of Millipore ultra pure integral 15 water purification system with A10 UV

An Annual service, calibration and validation of our clients ultra pure laboratory water system Millipore integral 15 with A10 uv Removal and replacement of all consumables supplied by Cryomed scientific cheaper than Millipore , replacement of RO membranes complete sanatisation of tank , all completed in one visit and customers system back fully operational operating… Continue reading Annual service,calibration and validation of Millipore ultra pure integral 15 water purification system with A10 UV

Annual service and Calibration of Leica cryostat

Customer contacted Cryomed scientific when their cryostat started showing ‘help’ on display, after servicing , calibration and reset of the help display cryostat is now back fully operational and operating at -25

Triple red’s SG ultra pure laboratory water system annual service and calibration to 18.2 m /ohms

Annual service and calibration of our customers Triple red / SG Ultra pure laboratory water purification system calibrated to 18.2 m/ohms, sanitisation of water storage tank and flushed to drain in the same day , fully back operational for our customer to use

Elga laboratory water purification system annual service and calibration to 18.2 m/ohms

The annual service and calibration of our customers Elga centra 200 laboratory ring main water purification system, consumables all changed sanitised and flushed to drain all within the same day , fully operational again for our customer to use with water quality at 18.2 m/ohms

-86 Ultra Low Temperature freezer servicing

An annual Panasonic and Gram bioline -80 freezer servicing and calibration, mobile ult freezer .

Avidity water system repair

Customer called Cryomed scientific when the manufacturer was unable to locate and repair water leak on the pure water system, within the hour our engineer located and replaced leaking part, System back fully operational and customer has asked for quotation to contract Cryomed scientific with all their water purification systems.

Annual -80 freezer servicing

Annual -80 freezer serving and calibration on these Eppendorf and Panasonic upright -80 freezers for our customer in Newcastle university

CBS PANASONIC LN2 VALIDATION

Cryomed scientific was contracted to perform the installation qualification and operational validation for their liquid nitrogen storage facility bank , 4 Panasonic CBS liquid nitrogen storage vessels validated and ready for performance validation in two days.

Oil separator removed and replaced on 2nd stage system of Thermo -80 ult freezer

New customer asked Cryomed scientific to diagnose the intermittent fault on their Thermo -80 upright ult freezer, freezer operating at -80 on arrival but after looking at temperature graphs it was easy to diagnose the intermittent failure of the 2nd stage refrigeration systems oil separator, Post ordering a replacement oil separator our engineer removed faulty… Continue reading Oil separator removed and replaced on 2nd stage system of Thermo -80 ult freezer

Annual service and calibration of laboratory fridges and freezers

Pleased to perform another annual service and calibration to our long standing client in edinburgh, Fridges and freezers serviced and single point calibration to traceable UKAS standards

-80 ULTRA LOW TEMPERATURE PANASONIC AND EPPENDORF FREEZER SERVICING AND CALIBRATION

Annual calibration and servicing of 2 -86 ult freezers for our client , Panasonic ult freezer and Eppendorf ult freezer, full calibration of the 20 comark temperature sensors of their monitoring system

Avidity Laboratory water system service and calibration

Annual service and calibration of Avidity ultra pure laboratory water system, consumables all replaced and timers reset, back in operation within a couple of hours, calibration completed to national standards

Coldroom repair

Customer contacted Cryomed scientific when their +4 walk-in coldroom failed, our engineer quickly diagnosed the system to have lost all its refrigerant charge, following a pressure test of the pipework a leak was found and repaired on the expansion valve flare nut, after vacuuming the system and fully charging the system with refrigerant the customer… Continue reading Coldroom repair

VINDON Climatic chamber repair

Customer contacted Cryomed scientific when their maintenance and validation company couldn’t repair their Vindon climatic chamber, Leak and pressure test of the refrigeration system identified a leak on the suction valve once the valve was fully tightened and repaired another leak and pressure test followed, fully charging the system with refrigerant, once the system fully… Continue reading VINDON Climatic chamber repair

Panasonic -80 ULT FREEZER COMPRESSOR REPLACEMENT

Removal and replacement of 2nd stage compressor on our customers Panasonic -80 upright ULT Freezer, Repair performed on site and fully operational by the end of the day, pressure tested, leak tested and fully charged with 1st and 2nd stage refrigerants, Saves customer time and money moving freezer off site,

Eppendorf Cryocube -80 upright freezer repair

New condenser fan replacement on Eppendorf Cryocube -80 upright freezer

Panasonic co2 incubator fault

Customer contacted Cryomed scientific to investigate co2 leaking from incubator and alarming, Attended site same day and our engineer found the incubator co2 inlet valve leaking , having a valve in stock we replaced valve and filters to have the incubator back fully operational within the hour

Elga water system service and validation

Annual service and validation of our customers Elga laboratory water purification system , consumables replaced and sanitisation of system, back in operation the same day.

Out of hours callout to Eppendorf Cryocube -80 Freezer

Customer calls Cryomed scientific 24 hour emergency callout hotline to attend their failed freezer at 10.00pm, a quick diagnostic of the fault by our engineer of the fan failure, temporary fan installed to keep freezer fully operational at -80 until new fan ordered, happy customer with samples all safe and secure.

Sanyo -80 ultra low temperature freezer fault

A new customer contacted Cryomed scientific to diagnose a fault with their Sanyo -80 ultra low temperature freezer, our engineer found the 2nd stage compressor seized and blowing fuse on start up. Quotation send to customer to perform the repair on site and return their freezer to -80

Thermo Revco ultra low -140 chest freezer service

Annual service of our clients Thermo Revco -140 freezer

Panasonic -80 upright ult freezer replacement 2nd stage compressor and oil sep

Replacement of 2nd stage compressor and oil separator on our clients Panasonic -80 ult freezer, total job performed and completed on customers premises ,our competitors usually remove from site to repair costing the end user more expense to have uplifted and delivered back and to be out of use for weeks instead of days,

Panasonic -86 ULT FREEZER FAULT DIAGNOSE

 Customer contacted Cryomed scientific to investigate intermittent issue with their Panasonic ultra low temperature freezer. The customer reported intermittent fault of temperature rising then dropping back down to temperature on a daily basis, Eventually our engineer traced the fault to the 2nd stage refrigeration system, new compressor and oil separator required, quote and go… Continue reading Panasonic -86 ULT FREEZER FAULT DIAGNOSE

Millipore essential 24 laboratory water purification system RO membrane and boost pump replacement

Customer in Southampton contracted Cryomed scientific to replace RO membranes and boost pump on their Millipore essential 24 water purification system, their operating quality is now 3m/s instead of 60m/s and now delivering 25 litres/hr instead of 6 litres/hr

Fisher Accu20 annual service

Nottingham university invited Cryomed to service their Fisher Accu20 water purification system, replaced all consumables and flushed system through, left system fully operational with water quality at 18.2 m/ohms

Labcold cooled incubator repair

Customer complained that the system chart was irratic in temperature recording and called Cryomed in to investigate, we found that the refrigeration system to have lost some of its refrigerant charge due to a leak on the charging schrader, we replaced schrader and vacuumed system, fully charged with refrigerant and now chart is operating at… Continue reading Labcold cooled incubator repair

Thermo Haake -50 dip cooler repair

Customer contacted Cryomed to repair their Haake Thermo -50 dip cooler , after leak tested both refrigeration systems , leaks were found and repaired, both systems vacuumed and re charged with relevant refrigerants , running the system down to -50 , cascade refrigeration system working perfectly

Edwards freeze dryer refrigeration repair

Customer in Edinburgh University contacted Cryomed to repair the refrigeration system on their 30 year old Edwards freeze dryer, refrigeration system required retrofitted to operate on a more suitable refrigerant to achieve -40, delighted customer when we achieved -42 under vacuum, another piece of equipment saved from being made redundant.

Thermo forma ult -86 upright freezer repair

New Customer contact Cryomed to investigate Thermo -80 freezer tripping circuit breaker , investigated issue and found faulty wiring tripping breaker on -80 Thermo freezer, repaired wiring and ult -80 freezer back down to set point of -85 in one visit, Happy customer that didn’t need to move any of their samples and reduce down… Continue reading Thermo forma ult -86 upright freezer repair

Elga laboratory pure water ring main system service

Annual service and calibration of this Elga laboratory pure water ring main system, saving customer over £1,000 on manufacturers costs for service and water consumable filters

Panasonic -80 ult chest freezer repair, control board replacement

Customer contacted Cryomed when over the weekend their Panasonic -80 ult freezer failed, we attended site and identified that the main control board was not sending a signal out to run 2nd stage compressor, we removed and replaced control board on the Panasonic -80 chest freezer , freezer was back down to -80 by the… Continue reading Panasonic -80 ult chest freezer repair, control board replacement

Panasonic co2 incubator gas leakage

Advised by customer hissing sound from Panasonic co2 incubator and losing co2 from bottle Leak tested system and found inlet co2 valve leaking, Due to Cryomed carrying spares for Panasonic co2 incubators we managed to remove and replace co2 valve in one visit and have the Panasonic incubator back fully operational for our happy customer.

-80 freezer repair and fan motor replacement

Customer requested fan motor changes to their -80 ult freezers, Fans replaced and alarm lights reset, no need for moving samples, freezer temperatures kept -80 throughout repairs, happy customer with a saving of over £1,000 on manufacturers repair.

Thermo forma co2 incubator service and repair

Customer contacted Cryomed to service and repair 3 of 4 Thermo co2 incubators, due to being over 20 years old we had to use parts from incubators to make 2 out of three operational, full calibration of temperature, co2 and humidity, new hepa filters , co2 and air filters installed

Eppendorf cooled centrifuge repair

IMG_0538 Customer reported Eppendorf cooled centrifuge not cooling, Identified leak on system, repaired on site, fully charged system with refrigerant and oil leaking the centrifuge fully operational on the first visit

New Brunswick -80 freezer 2nd stage compressor replacement

Customer reported New Brunswick scientific -80 freezer blowing fuses, Identified 2nd compressor faulty, removed and replaced compressor back up and running within 24hours and down to -80.

Validation and service of Purite Laboratory water system

Contacted by a new client to supply consumables, service and validate their Purite laboratory water purification system Service and validation completed with full documentation and consumables on time and saving our new client over £1,500 on their previous supplier

Sanyo ULT Freezer fan motor replacement

Customer reported temperature fluctuating on their Sanyo ultra low temperature freezer, identified the fault as a seized condenser fan motor , due to our high stock levels for all ult freezers we quickly removed and replaced fan motor in one visit leaving the freezer operating at -80.

Service ,calibration and condenser fan motor replacement

Removed and replaced condenser fan motors on Panasonic -150 freezers for Newcastle University customer

Service and replacement boost pump on Suez Purite laboratory water purification system

Newcastle university approached ourselves to service and replace boost pump on Purite water purification system when the manufacturer quoted , Cryomed scientific supplied the required consumables,serviced, calibrated temperature and water quality sensors and replace boost pump at half the cost quoted to the customer from Suez.

Avidity Pure water system service

University of Newcastle contacted Cryomed to supply consumables and service their Avidity water purification system , we supplied all the consumables ,serviced and calibrated quality sensors, saving the customer over 30% in cost than using the manufacturer.

Triple Red compact 2 laboratory water system service and repair

Cryomed were asked to repair and service triple red water system New consumables and reset of software and the system is back fully operational Delighted customer

UKAS CALIBRATION

Installataion qualification of new PHCBI Blood fridge and Biomedical freezer at blood bank, 3 point UKAS calibration of Labcold cooled incubator

Eppendorf 2nd stage compressor replacement

4 Thermo smart 2 pure laboratory water systems

  Service, calibration and validation of 4 Thermo smart 2 pure laboratory water systems, a massive cost saving for our new customer against manufacturer labour and consumables.

Sanyo -80 freezer repair

Millipore Elix100 service and calibration

Millipore laboratory water system service and validation including all consumables for Bristol Laboratories

Millipore Q integral 15 A10 laboratory water system service and validation

Eppendorf cooled centrifuge refrigeration compressor replacement

Haier ultra low temperature repair and service

RO membrane replacement to SG ultra pure water system

Elga Pure water system validation and service

Service and validation of a Elga ultra pure water system

Panasonic co2 incubators

Incubators over heating 2 control boards changed over , calibration of co2 and temperature levels on Panasonic co2 incubators Happy Customer ready continue their research

-20 Freezer Room condenser coil replacement

Removal and replacement of burst condenser coil on customers walk in -20 freezer room Room fully operational and back down to -20 set point in 3 hours

Ultra Pure Water System Installation and Validation

New ultra pure water system installed , commissioned and validated to clients requirements. Celtic Renewables are a new start up company and Cryomed Scientific are equipped to help with all their refrigeration and Ultra Pure water requirements.

Oil logging -86 Ultra low temperature freezer

Customer moved freezer to new location and never reached temperature in new location After 3 companies looking at freezer but unable to repair they called us in to investige problem. Found oil separator had failed internally and was flooding evaporator with oil then freezing at -40 We removed oil separator,cleaned out pipework with excess oil… Continue reading Oil logging -86 Ultra low temperature freezer