Annual service and calibration of our clients ultra low temperature -150 Panasonic freezer
- Laboratory Medical Equipment Service & Repairs
- 0131 358 9858
- 24hr Help Line - 07825 956815
- service@cryomedscientific.co.uk
Annual service and calibration of our clients ultra low temperature -150 Panasonic freezer
Annual service and calibration of Cryomed scientific’s client New Brunswick -80 upright freezers, our engineer identified 2 faulty alarm batteries and a faulty control board which was causing the compressors to short cycle intermittently, not an easy fault find.IMG_5282
Customer called Cryomed scientific when there was a noise coming from their Panasonic -150 freezer and an alarm was indicated, our engineer quickly diagnosed the fault as a faulty condenser fan motor which was removed and replaced within the hour, master reset of alarm timers and freezer was back fully operational and operating at -150… Continue reading PANASONIC -150 ULTRA LOW TEMPERATURE FREEZER FAN MOTOR REPLACEMENT AND TIMER RESET
A new customer contacted Cryomed scientific when their Eppendorf -80 freezer broke down , our engineer attended site and replaced the condenser fan motor on the 1st visit as we carry a fully comprehensive parts list for all -80 freezers, freezer fully operational and back down to -80 before the end of the day, ready… Continue reading Eppendorf replacement condenser fan motor
Removal and replacement of 2nd stage compressor on our customers Panasonic -80 upright ULT Freezer, Repair performed on site and fully operational by the end of the day, pressure tested, leak tested and fully charged with 1st and 2nd stage refrigerants, Saves customer time and money moving freezer off site,
Customer reported New Brunswick scientific -80 freezer blowing fuses, Identified 2nd compressor faulty, removed and replaced compressor back up and running within 24hours and down to -80.