Eppendorf co2 incubator Galaxy R co2 sensor replacement

Service, calibration and co2 sensor replacement to our clients Eppendorf co2 incubator, replaced ,calibrated and back fully operational

8 PHCBI CO2 INCUBATOR INSTALLATION

Installation of 8 new phcbi co2 incubators for our client, all completed and fully operational in one day.

PHCBI Co2 incubator with co2 leakage fault

Customer called Cryomed scientific when they discovered their phcbi co2 incubators using excessive amounts of co2 , on arrival our incubator specialist engineer quickly diagnosed the fault and rectified the faults and having the incubators fully operational on the first visit

Fisher Scientific Basic co2 Incubator installation and service

Installation and Calibration of a new Fisher basic co2 incubator for our client

Panasonic co2 incubator leak

Customer called Cryomed scientific when they realised they were using too much co2 and the co2 incubator wasn’t achieving set point of 5% Our engineer quickly diagnosed and located the faulty burst tubing in the back of the co2 incubator, repaired and customer was back operational within a couple of hours.

Panasonic co2 incubator fault

Customer contacted Cryomed scientific to investigate co2 leaking from incubator and alarming, Attended site same day and our engineer found the incubator co2 inlet valve leaking , having a valve in stock we replaced valve and filters to have the incubator back fully operational within the hour