Annual service of ult freezers and incubators

Annual service and calibration of our clients -150 Panasonic ult freezer, Haier ult freezer, co2 incubator and various fridges and freezers

Panasonic -150 chest freezer 2nd stage system refrigerant leak

IMG_7443IMG_7443 Refrigerant Leak on the 2nd stage of a Panasonic -150 ULT freezer model MDF-C2156VAN

Panasonic -150 Freezer leak and pressure test of 2nd stage system

A new client contacted Cryomed scientific when they found their Panasonic -150 ultra low freezer had failed, We uplifted the freezer back to our workshop when our engineer discovered the 2nd stage system of the Panasonic-150 freezer to have lost all its refrigerant charge. Once the Panasonic -150 freezer arrived at the workshop our engineers… Continue reading Panasonic -150 Freezer leak and pressure test of 2nd stage system

PANASONIC -150 ULTRA LOW TEMPERATURE FREEZER FAN MOTOR REPLACEMENT AND TIMER RESET

Customer called Cryomed scientific when there was a noise coming from their Panasonic -150 freezer and an alarm was indicated, our engineer quickly diagnosed the fault as a faulty condenser fan motor which was removed and replaced within the hour, master reset of alarm timers and freezer was back fully operational and operating at -150… Continue reading PANASONIC -150 ULTRA LOW TEMPERATURE FREEZER FAN MOTOR REPLACEMENT AND TIMER RESET