Thermo co2 incubators service and calibration

Annual service and calibration of our clients Thermo co2 incubators. Co2 and temperature calibrated by our certified geotech co2 and temperature analyser

Rs biotech galaxy r+ co2 sensor replacement and calibration

Client advised that their rs biotech co2 incubator levels were irratic, diagnosing the fault as the co2 sensor failure ,this was removed and replaced on the same visit, calibrated and back operational the same day, happy client.

Heraeus co2 incubator fan motor replacement

Customer contacted Cryomed Scientific when their Heraeus co2 incubator started fluctuating the co2 levels on the display and alarming, our engineer quickly assessed the issue as the internal fan motor seizure, after discussions with the department we then ordered and replaced the fan motor , fully calibrated co2 and temperature displays and left the incubator… Continue reading Heraeus co2 incubator fan motor replacement

Fisher Scientific Basic co2 Incubator installation and service

Installation and Calibration of a new Fisher basic co2 incubator for our client