Oil logging -86 Ultra low temperature freezer

Customer moved freezer to new location and never reached temperature in new location After 3 companies looking at freezer but unable to repair they called us in to investige problem. Found oil separator had failed internally and was flooding evaporator with oil then freezing at -40 We removed oil separator,cleaned out pipework with excess oil… Continue reading Oil logging -86 Ultra low temperature freezer