Heraeus co2 incubator fan motor replacement

Customer contacted Cryomed Scientific when their Heraeus co2 incubator started fluctuating the co2 levels on the display and alarming, our engineer quickly assessed the issue as the internal fan motor seizure, after discussions with the department we then ordered and replaced the fan motor , fully calibrated co2 and temperature displays and left the incubator… Continue reading Heraeus co2 incubator fan motor replacement

Nuaire co2 incubators repaired,serviced and calibrated

Customer decided to move from Avidity to Cryomed Scientific to repair, service and calibrate their co2 incubators. Removal and replacement of air pump cured their issue with irratic c02 levels on one incubator,both incubators had their air, co2 and hepa filters removed and replaced, calibration performed by our Gemtech analyser, happy customer and over £2,000… Continue reading Nuaire co2 incubators repaired,serviced and calibrated

Fisher Scientific Basic co2 Incubator installation and service

Installation and Calibration of a new Fisher basic co2 incubator for our client

BINDER CO2 INCUBATOR FAULTS

IMG_4947A new client of Cryomed scientific contracted our service and validation department to install and commission their newly acquired binder co2 incubator, once fully operational and found to have passed our onsite calibration it was found the temperature controller was faulty and causing the heater system to keep heating until the safety switch activated and… Continue reading BINDER CO2 INCUBATOR FAULTS

Service and calibration of Binder co2 and O2 controlled incubator

Annual service and calibration of Binder co2, oxygen level and temperature control levels