Heraeus co2 incubator fan motor replacement

Customer contacted Cryomed Scientific when their Heraeus co2 incubator started fluctuating the co2 levels on the display and alarming, our engineer quickly assessed the issue as the internal fan motor seizure, after discussions with the department we then ordered and replaced the fan motor , fully calibrated co2 and temperature displays and left the incubator… Continue reading Heraeus co2 incubator fan motor replacement

Nuaire co2 incubators repaired,serviced and calibrated

Customer decided to move from Avidity to Cryomed Scientific to repair, service and calibrate their co2 incubators. Removal and replacement of air pump cured their issue with irratic c02 levels on one incubator,both incubators had their air, co2 and hepa filters removed and replaced, calibration performed by our Gemtech analyser, happy customer and over £2,000… Continue reading Nuaire co2 incubators repaired,serviced and calibrated