Annual service, calibration and validation of Millipore ultra pure water system

Annual service, calibration and validation of our clients Millipore ultra pure laboratory water system, Cryomed supplied and replaced all consumables and calibrated the ultra pure water to 18.2 megaohms

Ultra pure water system service and validation

Annual service, calibration and validation of our clients triple red laboratory ultra pure water system.

Panasonic blood fridge service and calibration

6 monthly service and calibration of our clients Panasonic blood fridge chart recorder and temperature display of upper and lower temperature readings

Helmer cooled incubator service and calibration

6 monthly service and calibration of our clients Helmer cooled incubator

Eppendorf laboratory co2 incubator fault

Client reported to Cryomed scientific an issue with their Eppendorf co2 incubator, Cryomed scientifics engineer quickly diagnosed the fault as the incubator control sensor, once the new temperature sensor was received from Eppendorf the sensor was removed and replaced, calibrated and the Eppendorf co2 incubator was back fully operational

Rephile RO Water purification system installation

Cryomed scientific supplied and installed a new Raphile laboratory RO water system to our clients laboratory, installed, calibrated and fully operational in 4 hours

Elga laboratory ultra pure water system service

New client contacted Cryomed scientific when their Elga ultra pure laboratory water system wasn’t achieving 18.2 mega ohms, our engineer quickly diagnosed that the RO membranes required replacing and the system required a full service and calibration , RO membranes replaced and annual service of the Elga ultra pure water system left the system fully… Continue reading Elga laboratory ultra pure water system service

Hot Room over temperature issue

Our client called Cryomed scientific to investigate an issue with over heating on a constant temperature hot room, on investigation of the temperature controller our engineer found the controller to be faulty and operating the heaters at 2 degrees above set point, After close discussions with our client our engineer removed and replaced the temperature… Continue reading Hot Room over temperature issue

New Brunswick Scientific ULT -80 freezer servicing and calibration

Annual service and calibration of Cryomed scientific’s client New Brunswick -80 upright freezers, our engineer identified 2 faulty alarm batteries and a faulty control board which was causing the compressors to short cycle intermittently, not an easy fault find.IMG_5282

Ultra low temperature -80 freezers serviced and calibrated

Annual service and calibration of our clients 13 ultra low temperature -80 freezers

Thermo scientific replacement ultra pure filter on our clients laboratory water purification system

BINDER CO2 INCUBATOR FAULTS

IMG_4947A new client of Cryomed scientific contracted our service and validation department to install and commission their newly acquired binder co2 incubator, once fully operational and found to have passed our onsite calibration it was found the temperature controller was faulty and causing the heater system to keep heating until the safety switch activated and… Continue reading BINDER CO2 INCUBATOR FAULTS

Panasonic Blood fridge service and calibration

Avidity water purification service and calibration

Annual service, calibration and validation of our customers Avidity ultra pure laboratory water purification system, all consumables supplied by Cryomed scientific, 18.2 m/ohms calibration completed with our calibration meter, temperature and time calibrated with our UKAS certified equipment.

Service and calibration of Binder co2 and O2 controlled incubator

Annual service and calibration of Binder co2, oxygen level and temperature control levels

RS BIOTECH EX DEMO ULT FREEZER DELIVERY

Another delivery and installation of our stock ult freezers, This one is an rs biotech badged New Brunswick-80 upright freezer that was purchased by Newcastle university just before the Christmas break, All our ex-demo and refurbished-80 freezers come complete with 12 month parts and labour warranty, fully serviced, calibration certification and temperature mapping if requested.… Continue reading RS BIOTECH EX DEMO ULT FREEZER DELIVERY

Annual service,calibration and validation of Millipore ultra pure integral 15 water purification system with A10 UV

An Annual service, calibration and validation of our clients ultra pure laboratory water system Millipore integral 15 with A10 uv Removal and replacement of all consumables supplied by Cryomed scientific cheaper than Millipore , replacement of RO membranes complete sanatisation of tank , all completed in one visit and customers system back fully operational operating… Continue reading Annual service,calibration and validation of Millipore ultra pure integral 15 water purification system with A10 UV

Annual service and Calibration of Leica cryostat

Customer contacted Cryomed scientific when their cryostat started showing ‘help’ on display, after servicing , calibration and reset of the help display cryostat is now back fully operational and operating at -25

Annual service and calibration of laboratory fridges and freezers

Pleased to perform another annual service and calibration to our long standing client in edinburgh, Fridges and freezers serviced and single point calibration to traceable UKAS standards

-80 ULTRA LOW TEMPERATURE PANASONIC AND EPPENDORF FREEZER SERVICING AND CALIBRATION

Annual calibration and servicing of 2 -86 ult freezers for our client , Panasonic ult freezer and Eppendorf ult freezer, full calibration of the 20 comark temperature sensors of their monitoring system

Avidity Laboratory water system service and calibration

Annual service and calibration of Avidity ultra pure laboratory water system, consumables all replaced and timers reset, back in operation within a couple of hours, calibration completed to national standards

Elga water system service and validation

Annual service and validation of our customers Elga laboratory water purification system , consumables replaced and sanitisation of system, back in operation the same day.

Service ,calibration and condenser fan motor replacement

Removed and replaced condenser fan motors on Panasonic -150 freezers for Newcastle University customer

Service and replacement boost pump on Suez Purite laboratory water purification system

Newcastle university approached ourselves to service and replace boost pump on Purite water purification system when the manufacturer quoted , Cryomed scientific supplied the required consumables,serviced, calibrated temperature and water quality sensors and replace boost pump at half the cost quoted to the customer from Suez.

Millipore Elix100 service and calibration