Panasonic -150 Freezer leak and pressure test of 2nd stage system

A new client contacted Cryomed scientific when they found their Panasonic -150 ultra low freezer had failed, We uplifted the freezer back to our workshop when our engineer discovered the 2nd stage system of the Panasonic-150 freezer to have lost all its refrigerant charge. Once the Panasonic -150 freezer arrived at the workshop our engineers… Continue reading Panasonic -150 Freezer leak and pressure test of 2nd stage system

Oil separator removed and replaced on 2nd stage system of Thermo -80 ult freezer

New customer asked Cryomed scientific to diagnose the intermittent fault on their Thermo -80 upright ult freezer, freezer operating at -80 on arrival but after looking at temperature graphs it was easy to diagnose the intermittent failure of the 2nd stage refrigeration systems oil separator, Post ordering a replacement oil separator our engineer removed faulty… Continue reading Oil separator removed and replaced on 2nd stage system of Thermo -80 ult freezer

Panasonic -80 ULT FREEZER COMPRESSOR REPLACEMENT

Removal and replacement of 2nd stage compressor on our customers Panasonic -80 upright ULT Freezer, Repair performed on site and fully operational by the end of the day, pressure tested, leak tested and fully charged with 1st and 2nd stage refrigerants, Saves customer time and money moving freezer off site,