Infors shaker refrigerated incubators

Repairs of our customers refrigeration system on their infors refrigerated shaker incubator , also our engineer replaced the platelet actuator arms

Panasonic co2 incubator leak

Customer called Cryomed scientific when they realised they were using too much co2 and the co2 incubator wasn’t achieving set point of 5% Our engineer quickly diagnosed and located the faulty burst tubing in the back of the co2 incubator, repaired and customer was back operational within a couple of hours.

Panasonic co2 incubator fault

Customer contacted Cryomed scientific to investigate co2 leaking from incubator and alarming, Attended site same day and our engineer found the incubator co2 inlet valve leaking , having a valve in stock we replaced valve and filters to have the incubator back fully operational within the hour