Panasonic co2 incubator leak

Customer called Cryomed scientific when they realised they were using too much co2 and the co2 incubator wasn’t achieving set point of 5% Our engineer quickly diagnosed and located the faulty burst tubing in the back of the co2 incubator, repaired and customer was back operational within a couple of hours.

Infors cooled incubator shaker repair

Removal and replacement of internal air circulating fan motors and control board on our clients infors cooled incubator shaker system

BINDER CO2 INCUBATOR FAULTS

IMG_4947A new client of Cryomed scientific contracted our service and validation department to install and commission their newly acquired binder co2 incubator, once fully operational and found to have passed our onsite calibration it was found the temperature controller was faulty and causing the heater system to keep heating until the safety switch activated and… Continue reading BINDER CO2 INCUBATOR FAULTS

Panasonic co2 incubator fault

Customer contacted Cryomed scientific to investigate co2 leaking from incubator and alarming, Attended site same day and our engineer found the incubator co2 inlet valve leaking , having a valve in stock we replaced valve and filters to have the incubator back fully operational within the hour