PANASONIC -150 ULTRA LOW TEMPERATURE FREEZER FAN MOTOR REPLACEMENT AND TIMER RESET

Customer called Cryomed scientific when there was a noise coming from their Panasonic -150 freezer and an alarm was indicated, our engineer quickly diagnosed the fault as a faulty condenser fan motor which was removed and replaced within the hour, master reset of alarm timers and freezer was back fully operational and operating at -150… Continue reading PANASONIC -150 ULTRA LOW TEMPERATURE FREEZER FAN MOTOR REPLACEMENT AND TIMER RESET

Fan motor replacement on Heraeus -80 chest freezer

Customer called Cryomed scientific due to thei Heraeus ultra low temperature freezer overheating, our engineer quickly identified the condenser fan motor to be faulty, due to our large stock of ult freezer parts our engineer replaced the fan motor and had the freezer fully operational and back down to -80 within the hour.

Eppendorf replacement condenser fan motor

A new customer contacted Cryomed scientific when their Eppendorf -80 freezer broke down , our engineer attended site and replaced the condenser fan motor on the 1st visit as we carry a fully comprehensive parts list for all -80 freezers, freezer fully operational and back down to -80 before the end of the day, ready… Continue reading Eppendorf replacement condenser fan motor

VINDON Climatic chamber repair

Customer contacted Cryomed scientific when their maintenance and validation company couldn’t repair their Vindon climatic chamber, Leak and pressure test of the refrigeration system identified a leak on the suction valve once the valve was fully tightened and repaired another leak and pressure test followed, fully charging the system with refrigerant, once the system fully… Continue reading VINDON Climatic chamber repair

Eppendorf Cryocube -80 upright freezer repair

New condenser fan replacement on Eppendorf Cryocube -80 upright freezer

Sanyo ULT Freezer fan motor replacement

Customer reported temperature fluctuating on their Sanyo ultra low temperature freezer, identified the fault as a seized condenser fan motor , due to our high stock levels for all ult freezers we quickly removed and replaced fan motor in one visit leaving the freezer operating at -80.

Service ,calibration and condenser fan motor replacement

Removed and replaced condenser fan motors on Panasonic -150 freezers for Newcastle University customer