Nuaire co2 incubators repaired,serviced and calibrated

Customer decided to move from Avidity to Cryomed Scientific to repair, service and calibrate their co2 incubators. Removal and replacement of air pump cured their issue with irratic c02 levels on one incubator,both incubators had their air, co2 and hepa filters removed and replaced, calibration performed by our Gemtech analyser, happy customer and over £2,000… Continue reading Nuaire co2 incubators repaired,serviced and calibrated

Panasonic co2 incubator fault

Customer contacted Cryomed scientific to investigate co2 leaking from incubator and alarming, Attended site same day and our engineer found the incubator co2 inlet valve leaking , having a valve in stock we replaced valve and filters to have the incubator back fully operational within the hour