Triple Red Avidity Science ultra pure water system fault

client called Cryomed scientific to report quality issues with the triple red Avidity science ultra pure water system, our engineer quickly diagnosed the fault as the quality sensor who then removed and replaced the sensor, calibrated the water quality by using our ultra pure water quality meter and retuned system to service

Nuaire co2 incubators repaired,serviced and calibrated

Customer decided to move from Avidity to Cryomed Scientific to repair, service and calibrate their co2 incubators. Removal and replacement of air pump cured their issue with irratic c02 levels on one incubator,both incubators had their air, co2 and hepa filters removed and replaced, calibration performed by our Gemtech analyser, happy customer and over £2,000… Continue reading Nuaire co2 incubators repaired,serviced and calibrated

Avidity laboratory water system pump replacement

Customer contacted Cryomed scientific when they had a water leak from their Avidity science water purification system, our engineer quickly diagnosed the fault as the pump head leaking, pump was removed and replaced with system back fully operational.