New Brunswick Scientific ULT -80 freezer servicing and calibration

Annual service and calibration of Cryomed scientific’s client New Brunswick -80 upright freezers, our engineer identified 2 faulty alarm batteries and a faulty control board which was causing the compressors to short cycle intermittently, not an easy fault find.IMG_5282

SANYO ULT freezer 2nd stage system choking

Customer reported freezer fluctuating in temperature, on investigation by our ULT freezer engineer it was quickly diagnosed the 2nd stage refrigeration system was partially choking, removal and installation of a new filter drier on the 2nd stage refrigeration system cured the fault, freezer fully back operational and another happy customer.

Fan motor replacement on Heraeus -80 chest freezer

Customer called Cryomed scientific due to thei Heraeus ultra low temperature freezer overheating, our engineer quickly identified the condenser fan motor to be faulty, due to our large stock of ult freezer parts our engineer replaced the fan motor and had the freezer fully operational and back down to -80 within the hour.

Eppendorf replacement condenser fan motor

A new customer contacted Cryomed scientific when their Eppendorf -80 freezer broke down , our engineer attended site and replaced the condenser fan motor on the 1st visit as we carry a fully comprehensive parts list for all -80 freezers, freezer fully operational and back down to -80 before the end of the day, ready… Continue reading Eppendorf replacement condenser fan motor

RS BIOTECH EX DEMO ULT FREEZER DELIVERY

Another delivery and installation of our stock ult freezers, This one is an rs biotech badged New Brunswick-80 upright freezer that was purchased by Newcastle university just before the Christmas break, All our ex-demo and refurbished-80 freezers come complete with 12 month parts and labour warranty, fully serviced, calibration certification and temperature mapping if requested.… Continue reading RS BIOTECH EX DEMO ULT FREEZER DELIVERY

Panasonic -80 ULT FREEZER COMPRESSOR REPLACEMENT

Removal and replacement of 2nd stage compressor on our customers Panasonic -80 upright ULT Freezer, Repair performed on site and fully operational by the end of the day, pressure tested, leak tested and fully charged with 1st and 2nd stage refrigerants, Saves customer time and money moving freezer off site,

Out of hours callout to Eppendorf Cryocube -80 Freezer

Customer calls Cryomed scientific 24 hour emergency callout hotline to attend their failed freezer at 10.00pm, a quick diagnostic of the fault by our engineer of the fan failure, temporary fan installed to keep freezer fully operational at -80 until new fan ordered, happy customer with samples all safe and secure.

Panasonic -80 upright ult freezer replacement 2nd stage compressor and oil sep

Replacement of 2nd stage compressor and oil separator on our clients Panasonic -80 ult freezer, total job performed and completed on customers premises ,our competitors usually remove from site to repair costing the end user more expense to have uplifted and delivered back and to be out of use for weeks instead of days,

New Brunswick -80 freezer 2nd stage compressor replacement

Customer reported New Brunswick scientific -80 freezer blowing fuses, Identified 2nd compressor faulty, removed and replaced compressor back up and running within 24hours and down to -80.

Sanyo ULT Freezer fan motor replacement

Customer reported temperature fluctuating on their Sanyo ultra low temperature freezer, identified the fault as a seized condenser fan motor , due to our high stock levels for all ult freezers we quickly removed and replaced fan motor in one visit leaving the freezer operating at -80.