SANYO ULT freezer 2nd stage system choking

Customer reported freezer fluctuating in temperature, on investigation by our ULT freezer engineer it was quickly diagnosed the 2nd stage refrigeration system was partially choking, removal and installation of a new filter drier on the 2nd stage refrigeration system cured the fault, freezer fully back operational and another happy customer.

Sanyo -80 ultra low temperature freezer fault

A new customer contacted Cryomed scientific to diagnose a fault with their Sanyo -80 ultra low temperature freezer, our engineer found the 2nd stage compressor seized and blowing fuse on start up. Quotation send to customer to perform the repair on site and return their freezer to -80

Sanyo ULT Freezer fan motor replacement

Customer reported temperature fluctuating on their Sanyo ultra low temperature freezer, identified the fault as a seized condenser fan motor , due to our high stock levels for all ult freezers we quickly removed and replaced fan motor in one visit leaving the freezer operating at -80.

Panasonic co2 incubators

Incubators over heating 2 control boards changed over , calibration of co2 and temperature levels on Panasonic co2 incubators Happy Customer ready continue their research