Panasonic ULT freezer 2nd stage compressor replacement

Panasonic -80 freezer 2nd stage compressor removal and replacement, full vacuum and pressure test of system and back down to -80 and ready for use once again within 24 hours

Panasonic ult chest freezer compressor failure

SANYO ULT freezer 2nd stage system choking

Customer reported freezer fluctuating in temperature, on investigation by our ULT freezer engineer it was quickly diagnosed the 2nd stage refrigeration system was partially choking, removal and installation of a new filter drier on the 2nd stage refrigeration system cured the fault, freezer fully back operational and another happy customer.

-86 Ultra Low Temperature freezer servicing

An annual Panasonic and Gram bioline -80 freezer servicing and calibration, mobile ult freezer .

-80 ULTRA LOW TEMPERATURE PANASONIC AND EPPENDORF FREEZER SERVICING AND CALIBRATION