New customer asked Cryomed scientific to diagnose the intermittent fault on their Thermo -80 upright ult freezer, freezer operating at -80 on arrival but after looking at temperature graphs it was easy to diagnose the intermittent failure of the 2nd stage refrigeration systems oil separator, Post ordering a replacement oil separator our engineer removed faulty… Continue reading Oil separator removed and replaced on 2nd stage system of Thermo -80 ult freezer
Removal and replacement of 2nd stage compressor on our customers Panasonic -80 upright ULT Freezer, Repair performed on site and fully operational by the end of the day, pressure tested, leak tested and fully charged with 1st and 2nd stage refrigerants, Saves customer time and money moving freezer off site,
Replacement of 2nd stage compressor and oil separator on our clients Panasonic -80 ult freezer, total job performed and completed on customers premises ,our competitors usually remove from site to repair costing the end user more expense to have uplifted and delivered back and to be out of use for weeks instead of days,