Eppendorf -80 ult freezer door vacuum release issue

Customer called Cryomed scientific when they couldn’t open the outer freezer door to access their samples due to the door being jammed closed, Our engineer quickly diagnosed the issue as being an ice build up on the inner door over the vacuum release system, once the ice was removed the door was now easily opened… Continue reading Eppendorf -80 ult freezer door vacuum release issue

Purite ultra pure laboratory water system leak

IMG_5870 Client called Cryomed scientific when their Purite water purification system starting leaking , on closer investigation from our engineer he traced the leak to a faulty inlet water pressure guage, our water specialist quickly removed and replaced the leaking water pressure guage with a new one and the Purite ultra pure water system was… Continue reading Purite ultra pure laboratory water system leak

HAIER AND LABCOLD ULT -80 FREEZER SERVICING AND CALIBRATION

Annual service and calibration for our clients Haier and Labcold -80 ultra low temperature freezers

Millipore Ultra pure Laboratory water system fault

A call from our client regards flow issues with their Millipore laboratory ultra pure water system , our engineer quickly diagnosed the issue as being a failure of the flow meter on the Millipore system, removal and replacement of the flow meter quickly resolved the issue and the system was placed back into full operation.

Hot Room over temperature issue

Our client called Cryomed scientific to investigate an issue with over heating on a constant temperature hot room, on investigation of the temperature controller our engineer found the controller to be faulty and operating the heaters at 2 degrees above set point, After close discussions with our client our engineer removed and replaced the temperature… Continue reading Hot Room over temperature issue