VINDON Climatic chamber repair

Customer contacted Cryomed scientific when their maintenance and validation company couldn’t repair their Vindon climatic chamber, Leak and pressure test of the refrigeration system identified a leak on the suction valve once the valve was fully tightened and repaired another leak and pressure test followed, fully charging the system with refrigerant, once the system fully… Continue reading VINDON Climatic chamber repair

Panasonic -80 ULT FREEZER COMPRESSOR REPLACEMENT

Removal and replacement of 2nd stage compressor on our customers Panasonic -80 upright ULT Freezer, Repair performed on site and fully operational by the end of the day, pressure tested, leak tested and fully charged with 1st and 2nd stage refrigerants, Saves customer time and money moving freezer off site,