Panasonic -80 ult chest freezer repair, control board replacement

Customer contacted Cryomed when over the weekend their Panasonic -80 ult freezer failed, we attended site and identified that the main control board was not sending a signal out to run 2nd stage compressor, we removed and replaced control board on the Panasonic -80 chest freezer , freezer was back down to -80 by the… Continue reading Panasonic -80 ult chest freezer repair, control board replacement

Panasonic co2 incubator gas leakage

Advised by customer hissing sound from Panasonic co2 incubator and losing co2 from bottle Leak tested system and found inlet co2 valve leaking, Due to Cryomed carrying spares for Panasonic co2 incubators we managed to remove and replace co2 valve in one visit and have the Panasonic incubator back fully operational for our happy customer.